Viral RNA/DNA Extraction and Quantification

Procedure

Preparation of lysate:

  • Add 200 μL of sample to the microcentrifuge tube and add 25 μL Proteinase K.
  • Add 200 μL Lysis Buffer and mix by vortexing.
  • Incubate at 56 °C for 15 minutes.
  • Add 250 μL 100% ethanol to the tube and mix by vortexing.
  • Incubate the lysate for 5 minutes at room temperature.
  • Add the above lysate to the Viral Spin Column in a collection tube.
  • Centrifuge the column at 6800 × g for 1 minute. Discard the collection tube. Place the spin column in a new Wash Tube.
  • Wash the column with 500 μL Wash Buffer (WII) with ethanol. Centrifuge at 6800 × g for 1 minute. Discard the flow through.
  • Repeat the wash step.
  • Centrifuge the spin column at maximum speed for 1 minute to remove any residual Wash Buffer.
  • Elute with 10-50 μL nuclease-free water.
  • Incubate at room temperature for 1 minute. Centrifuge the spin column at maximum speed for 1 minute to elute nucleic acids.
  • To quantify the Viral RNA/DNA, blank the NanoDrop instrument using nuclease-free water. Load a small volume of the extracted viral RNA/DNA sample and measure absorbance.